# The Measuring Instruments Regulations 2016, Paragraph 2.4.1


For minimum measured quantities greater than or equal to 2 litres the following conditions apply:
Condition 1
Emin shall fulfil the condition: Emin ≥ 2 R, where R is the smallest scale interval of the indication device.
Condition 2
Emin is given by the formula: Emin = (2MMQ)x (A/100) where:
— MMQ is the minimum measured quantity,
— A is the numerical value specified in line A of Table 2.

---
Source: https://www.legislation.gov.uk/uksi/2016/1153/schedule/1F/paragraph/2.4.1 (the official version)
Page: https://openlegislation.uk/measuring-instruments-regulations-2016/schedule-1F/paragraph-2.4.1
Contains public sector information licensed under the Open Government Licence v3.0. Source: legislation.gov.uk, The National Archives.