# The Measuring Instruments Regulations 2016, Paragraph 7: Putting into use


Table 5
Accuracy class
Types of Measuring system
0.3
Measuring systems on pipeline
0.5
All measuring systems if not differently stated elsewhere in this Table, in particular:
fuel dispensers (not for liquefied gases),
measuring systems on road tankers for liquids of low viscosity (< 20 mPa.s)
1.0
Measuring systems for liquefied gases under pressure measured at a temperature equal to or above – 10 °C
Measuring systems normally in class 0.3 or 0.5 but used for liquids
whose temperature is less than – 10 °C or greater than 50 °C
whose dynamic viscosity is higher than 1,000 mPa.s
whose maximum volumetric flowrate is not higher than 20 L/h
1.5
Measuring systems for liquefied gases under pressure measured at a temperature below – 10 °C (other than cryogenic liquids)
2.5
Measuring systems for cryogenic liquids (temperature below – 153 °C)

---