# The Measuring Instruments Regulations 2016, Paragraph 2.3: Verification scale interval — multi-interval instruments


Table 3
1 For i = r the corresponding column of Table 2 applies with e replaced by er.
Verification scale interval
Number of verification scale intervalsn = Max/e
Minimum value1
n = Maxi /e(i+1)
For i = r the corresponding column of Table 2 applies with e replaced by er.
Maximum value
n = Maxi /ei
Y(I)
0.001 g ≤ ei
50,000
Y(II)
0.001 g ≤ ei≤ 0.05 g
5,000
100,000
0.1 g ≤ ei
5,000
100,000
Y(a)
0.1 g ≤ ei
500
10,000
Y(b)
5 g ≤ ei
50
1 000
Where:
i = 1, 2, … r
i = partial weighing range
r = total number of partial ranges

---
Source: https://www.legislation.gov.uk/uksi/2016/1153/schedule/1G/chapter/II/paragraph/2.3 (the official version)