# The Measuring Instruments Regulations 2016, Paragraph 5.1


The MPE due to an influence factor shall be as specified in Table 10.
Table 10
Load (m) in verification scale intervals (d)
MPE
0 < m ≤ 500
± 0.5 d
500 < m ≤ 2,000
± 1.0 d
2,000 < m ≤ 10,000
± 1.5 d

---