  "heading": "Other requirements",
  "text": "The resolution shall be equal to or of one order of magnitude higher than the values shown in Table 3.\nTable 3\nResolution\nCO\nCO2\nO2\nHC\nClass O and class I\n0.01 % vol\n0.1 % vol\n0.01 % vol for measurand values below or equal to 4 % vol, otherwise 0.1 % vol.\n1 ppm vol\nThe lambda value shall be displayed with a resolution of 0.001.\nThe standard deviation of 20 measurements shall not be greater than one third of the modulus of the MPE for each applicable gas volume fraction.",
  "extent": null,
  "status": null,
  "version": "current",
  "annotations": [
    {
      "type": "F",
      "label": "F1",
      "text": "Schs. 1A-1K inserted (E.W.S.) (31.12.2020) by The Product Safety and Metrology etc. (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/696), reg. 1, Sch. 27 para. 49 (with Sch. 27 para. 50(a)) (as amended by S.I. 2020/676, regs. 1(1), 2); 2020 c. 1, Sch. 5 para. 1(1)"
    }
  ],
  "outstanding_effects": [],
  "provenance": {
    "source": "legislation.gov.uk",
    "source_url": "https://www.legislation.gov.uk/uksi/2016/1153/schedule/1J/paragraph/6",
    "retrieved_at": "2026-10-28T23:17:40.726189+00:00"
  },
  "licence": "Open Government Licence v3.0",
  "licence_url": "https://www.nationalarchives.gov.uk/doc/open-government-licence/version/3/",
  "attribution": "Contains public sector information licensed under the Open Government Licence v3.0. Source: legislation.gov.uk, The National Archives."
}