# The Measuring Instruments Regulations 2016, Paragraph 6: Other requirements


The resolution shall be equal to or of one order of magnitude higher than the values shown in Table 3.
Table 3
Resolution
CO
CO2
O2
HC
Class O and class I
0.01 % vol
0.1 % vol
0.01 % vol for measurand values below or equal to 4 % vol, otherwise 0.1 % vol.
1 ppm vol
The lambda value shall be displayed with a resolution of 0.001.
The standard deviation of 20 measurements shall not be greater than one third of the modulus of the MPE for each applicable gas volume fraction.

---