  "heading": "Test loads",
  "text": "(1) Every repaired meter shall be tested at each of the loads specified in table 2, except that test number 2 in that table may be omitted.\n(2) Every new meter shall be calibrated and tested at a sufficient number of load points so as to ensure that meter errors are not greater than the limits specified in paragraph 9.\n(3) A test for starting shall also be carried out on new meters in accordance with test number 7 of table 2.\n(4) The ratio errors shall be determined for voltage transformers that are intended for use with meters but are not tested with a meter.\n(5) Current transformers intended for use with meters but not tested with a meter shall be tested from 5 per cent. to 120 per cent. of rated current.",
  "extent": "E+W+S",
  "status": null,
  "version": "current",
  "annotations": [
    {
      "type": "I",
      "label": "I1",
      "text": "Sch. 1 para. 6 in force at 1.8.1998, see reg. 1"
    }
  ],
  "outstanding_effects": [],
  "provenance": {
    "source": "legislation.gov.uk",
    "source_url": "https://www.legislation.gov.uk/uksi/1998/1566/schedule/1/paragraph/6",
    "retrieved_at": "2026-10-29T11:26:36.743610+00:00"
  },
  "licence": "Open Government Licence v3.0",
  "licence_url": "https://www.nationalarchives.gov.uk/doc/open-government-licence/version/3/",
  "attribution": "Contains public sector information licensed under the Open Government Licence v3.0. Source: legislation.gov.uk, The National Archives."
}