Paragraph 3 Units of mass

Original text. Later amendments are not shown.

Classification

3.1.

Instruments with one weighing range

Instruments equipped with an auxiliary indicating device shall belong to class I or class II. For these instruments the minimum capacity lower limits for these two classes are obtained from Table 1 by replacement in column 3 of the verification scale interval (e) by the actual scale interval (d).

If d < 10−4 g, the maximum capacity of class I may be less than 50 000 e.

3.2.

Instruments with multiple weighing ranges

Multiple weighing ranges are permitted, provided they are clearly indicated on the instrument. Each individual weighing range is classified according to 3.1. If the weighing ranges fall into different accuracy classes the instruments shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.

3.3.

Multi-interval instruments

3.3.1.

Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).

Multi-interval instruments shall not be equipped what an auxiliary indicating device.

3.3.2.

Each partial weighing range i of multi-interval instruments is defined by:

— its verification scale interval eiwith e*(i+1) > ei
— its maximum capacity Maxiwith Maxi = Max
— its minimum capacity Miniwith Mini = Max( − 1)
and Min1 = Min

where:

  • i = 1, 2, . . . r,

  • i = partial weighing range number,

  • r = the total number of partial weighing ranges.

All capacities are capacities of net load, irrespective of the value of any tare used.

3.3.3.

The partial weighing ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, this class being the instrument’s accuracy class.

Table 2

Multi-interval instruments

ClassVerification scale interval (e)Minimum capacity (Min)Number of verification scale intervals
Minimum valueMinimum valuef1Maximum value
n=Maxie(i+1) n=Maxiei
(1)

For i = r the corresponding column of Table 1 applies, with e replaced by er.

I0,001 g ≤ ei100 ei50 000—
II0,001 g ≤ ei ≤ 0,05 g20 ei5 000100 000
0,1 g ≤ ei50 ei5 000100 000
III0,1 g ≤ ei20 ei50010 000
IIII5 g ≤ ei10 ei501 000