# The Non-automatic Weighing Instruments Regulations 2016, Paragraph 3.3.2


Each partial weighing range i of multi-interval instruments is defined by—
— its verification scale interval ei
with e(i + 1) > ei
— its maximum capacity Maxi
with Maxr = Max
— its minimum capacity Mini
with Mini = Max (i – 1)
and Min1 = Min
Where:
i = 1, 2, …r,
i = partial weighing range number,
r = the total number of partial weighing ranges
All capacities are capacities of net load, irrespective of the value of any tare used.

---