# The Non-Road Mobile Machinery (Emission of Gaseous and Particulate Pollutants) Regulations 1999, Appendix 2


1 CALIBRATION OF THE ANALYTICAL INSTRUMENTS
Introduction
1.1 Each analyser shall be calibrated as often as necessary to fulfil the accuracy requirements of this standard. The calibration method that shall be used is described in this paragraph for the analysers indicated in Appendix 1, section 1.4.3.
Calibration gases
1.2 The shelf life of all calibration gases must be respected.
The expiry date of the calibration gases stated by the manufacturer shall be recorded.
Pure gases
1.2.1 The required purity of the gases is defined by the contamination limits given below. The following gases must be available for operation:
purified nitrogen
(contamination ≤ 1 ppm C, ≤ 1 ppm CO, ≤ 400 ppm CO2, ≤ 0.1 ppm NO)
purified oxygen
(purity > 99.5% vol 02)
hydrogen-helium mixture
(40 ± 2% hydrogen, balance helium)
(contamination ≤ 1 ppm C, ≤ 400 ppm CO)
purified synthetic air
(contamination ≤ 1 ppm C, ≤ 1 ppm CO, ≤ 400 ppm CO2,
≤ 0.1 ppm NO)
(oxygen content between 18-21% vol)
Calibration and span gases
1.2.2 Mixture of gases having the following chemical compositions shall be available:
C3H8 and purified synthetic air (see section 1.2.1)
CO and purified nitrogen
NO and purified nitrogen (the amount of NO2 contained in this calibration gas must not exceed 5% of the NO content)
O2 and purified nitrogen
CO2 and purified nitrogen
CH4 and purified synthetic air
C2H6 and purified synthetic air
Note: other gas combinations are allowed provided the gases do not react with one another.
The true concentration of a calibration and span gas must be within ± 2% of the nominal value. All concentrations of calibration gas shall be given on a volume basis (volume percent or volume ppm).
The gases used for calibration and span may also be obtained by means of a gas divider, diluting with purified N2 or with purified synthetic air. The accuracy of the mixing device must be such that the concentration of the diluted calibration gases may be determined to within ±2%.
Operating procedure for analysers and sampling system
1.3 The operating procedure for analysers shall follow the start-up and operating instructions of the instrument manufacturer. The minimum requirements given in sections 1.4 to 1.9 shall be included.
Leakage test
1.4 A system leakage test shall be performed. The probe shall be disconnected from the exhaust system and the end plugged. The analyser pump shall be switched on. After an initial stabilisation period all flow meters should read zero. If not, the sampling lines shall be checked and the fault corrected. The maximum allowable leakage rate on the vacuum side shall be 0.5% of the in-use flow rate for the portion of the system being checked. The analyser flows and bypass flows may be used to estimate the in-use flow rates.
Another method is the introduction of a concentration step change at the beginning of the sampling line by switching from zero to span gas.
If after an adequate period of time the reading shows a lower concentration compared to the introduced concentration, this points to calibration or leakage problems.
1.5 Calibration procedure
Instrument assembly
1.5.1 The instrument assembly shall be calibrated and calibration curves checked against standard gases. The same gas flow rates shall be used as when sampling exhaust.
Warming-up time
1.5.2 The warming-up time should be according to the recommendations of the manufacturer. If not specified, a minimum of two hours is recommended for warming-up the analysers.
NDIR and HFID analyser
1.5.3 The NDIR analyser shall be tuned, as necessary, and the combustion flame of the HFID analyser shall be optimized (section 1.8.1).
Calibration
1.5.4 Each normally used operating range shall be calibrated.
Using purified synthetic air (or nitrogen), the CO, CO2, NOX, HC and O2 analysers shall be set at zero.
The appropriate calibration gases shall be introduced to the analysers, the values recorded, and the calibration curve established according to section 1.5.6.
The zero setting shall be re-checked and the calibration procedure repeated, if necessary.
1.5.5 Establishment of the calibration curve
General guidelines
1.5.5.1 The analyser calibration curve is established by at least five calibration points (excluding zero) spaced as uniformly as possible. The highest nominal concentration must be equal to or higher than 90% of full scale.
The calibration curve is calculated by the method of least squares. If the resulting polynomial degree is greater than three, the number of calibration points (zero included) must be at least equal to this polynomial degree plus two.
The calibration curve must not differ by more than ±2% from the nominal value of each calibration point and by more than ±1% of full scale at zero
From the calibration curve and the calibration points, it is possible to verify that the calibration has been carried out correctly. The different characteristic parameters of the analyser must be indicated, particularly:
the measuring range,
the sensitivity,
the date of carrying out the calibration.
Calibration below 15% of full scale
1.5.5.2 The analyser calibration curve is established by at least ten calibration points (excluding zero) spaced so that 50% of the calibration points are below 10% of full scale.
The calibration curve is calculated by the method of least squares.
The calibration curve must not differ by more than ± 4% from the nominal value of each calibration point and by more than ± 1% of full scale at zero.
Alternative methods
1.5.5.3 If it can be shown that alternative technology (eg computer, electronically controlled range switch, etc) can give equivalent accuracy, then these alternatives may be used.
Verification of the calibration
1.6 Each normally used operating range shall be checked prior to each analysis in accordance with the following procedure.
The calibration is checked by using a zero gas and a span gas whose nominal value is more than 80% of full scale of the measuring range.
If, for the two points considered, the value found does not differ by more than ± 4% of full scale from the declared reference value, the adjustment parameters may be modified. Should this not be the case, a new calibration curve shall be established in accordance with section 1.5.4.
Efficiency test of the NOX converter
1.7 The efficiency of the converter used for the conversion of NO2 into NO is tested as given in sections 1.7.1 to 1.7.8 (Figure 1).
Test set-up
1.7.1 Using the test set-up as shown in Figure 1 (see also Appendix 1, section 1.4.3.5) and the procedure below, the efficiency of converters can be tested by means of an ozonator.
Figure 1
Schematic of NO2 converter efficiency device
Calibration
1.7.2 The CLD and the HCLD shall be calibrated in the most common operating range following the manufacturer’s specifications using zero and span gas (the NO content of which must amount to about 80% of the operating range and the NO2 concentration of the gas mixture to less than 5% of the NO concentration). The NOX analyser must be in the NO mode so that the span gas does not pass through the converter. The indicated concentration has to be recorded.
Calculation
1.7.3 The efficiency of the NOX converter is calculated as follows:
Efficiency (%)={1+a-bc-d}x100 (a) NOX concentration according to section 1.7.6;
(b) NOX concentration according to section 1.7.7;
(c) NO concentration according to section 1.7.4;.
(d) NO concentration according to section 1.7.5.
Adding of oxygen
1.7.4 Via a T-fitting, oxygen or zero air is added continuously to the gas flow until the concentration indicated is about 20% less than the indicated calibration concentration given in section 1.7.2. (The analyser is in the NO mode).
The indicated concentration (c) shall be recorded. The ozonator is kept de-activated throughout the process.
Activation of the ozonator
1.7.5 The ozonator is now activated to generate enough ozone to bring the NO concentration down to about 20% (minimum 10%) of the calibration concentration given in section 1.7.2. The indicated concentration (d) shall be recorded. (The analyser is in the NO mode.)
NOX mode
1.7.6 The NO analyser is then switched to NOX mode so that the gas mixture (consisting of NO, NO2, O2 and N2) now passes through the converter. The indicated concentration (a) shall be recorded. (The analyser is in the NOX mode.)
De-activation of the ozonator
1.7.7 The ozonator is now de-activated. The mixture of gases described in section 1.7.6 passes through the converter into the detector. The indicated concentration (b) shall be recorded. (The analyser is the NOX mode.)
NO mode
1.7.8 Switched to NO mode with the ozonator de-activated, the flow of oxygen or synthetic air is also shut off. The NOX reading of the analyser shall not deviate by more than ± 5% from the value measured according to section 1.7.2. (The analyser is in the NO mode.)
Test interval
1.7.9 The efficiency of the converter must be tested prior to each calibration of the NOX analyser.
Efficiency requirement
1.7.10 The efficiency of the converter shall not be less than 90%, but a higher efficiency of 95% is strongly recommended.
Note: If, with the analyser in the most common range, the ozonator cannot give a reduction from 80% to 20% according to section 1.7.5, then the highest range which will give the reduction shall be used.
1.8 Adjustment of the FID
Optimization of the detector response
1.8.1 The HFID must be adjusted as specified by the instrument manufacturer. A propane in air span gas should be used to optimize the response on the most common operating range.
With the fuel and air flow rates set at the manufacturer’s recommendations, a 350 ± 75 ppm C span gas shall be introduced to the analyser. The response at a given fuel flow shall be determined from the difference between the span gas response and the zero gas response. The fuel flow shall be incrementally adjusted above and below the manufacturer’s specification. The span and zero response at these fuel flows shall be recorded. The difference between the span and zero response shall be plotted and the fuel flow adjusted to the rich side of the curve.
Hydrocarbon response factors
1.8.2 The analyser shall be calibrated using propane in air and purified synthetic air, according to section 1.5.
Response factors shall be determined when introducing an analyser into service and after major service intervals. The response factor (Rf) for a particular hydrocarbon species is the ratio of the FID C1 reading to the gas concentration in the cylinder expressed by ppm C1.
The concentration of the test gas must be at a level to give a response of approximately 80% of full scale. The concentration must be known to an accuracy of ± 2% in reference to a gravimetric standard expressed in volume. In addition, the gas cylinder must be pre-conditioned for 24 hours at a temperature of 298 K (25°C) ± 5 K.
The test gases to be used and the recommended relative response factor ranges are as follows:
— methane and purified synthetic air:
1.00 ≤ Rf ≤ 1.15
— propylene and purified synthetic air:
0.90 ≤ Rf ≤ 1.1
— toluene and purified synthetic air:
0.90 ≤ Rf ≤ 1.10
These values are relative to the response factor (Rf) of 1.00 for propane and purified synthetic air.
Oxygen interference check
1.8.3 The oxygen interference check shall be determined when introducing an analyser into service and after major service intervals.
The response factor is defined and shall be determined as described in section 1.8.2. The test gas to be used and the recommended relative response factor range are as follows:
—propane and nitrogen:
0.95 ≤ Rf ≤ 1.05
This value is relative to the response factor (Rf) of 1.00 for propane and purified synthetic air.
The FID burner air oxygen concentration must be within ± 1 mole % of the oxygen concentration of the burner air used in the latest oxygen interference check. If the difference is greater, the oxygen interference must be checked and the analyser adjusted, if necessary.
Interference effects with NDIR and CLD analysers
1.9 Gases present in the exhaust other than the one being analysed can interfere with the reading in several ways. Positive interference occurs in NDIR instruments where the interfering gas gives the same effect as the gas being measured, but to a lesser degree. Negative interference occurs in NDIR instruments by the interfering gas broadening the absorption band of the measured gas, and in CLD instruments by the interfering gas quenching the radiation. The interference checks in sections 1.9.1 and 1.9.2 shall be performed prior to an analyser’s initial use and after major service intervals.
CO analyser interference check
1.9.1 Water and CO2 can interfere with the CO analyser performance. Therefore a CO2 span gas having a concentration of 80 to 100% of full scale of the maximum operating range used during testing shall be bubbled through water at room temperature and the analyser response recorded. The analyser response must not be more than 1% of full scale for ranges equal to or above 300 ppm or more than 3 ppm for ranges below 300 ppm.
NOX analyser quench checks
1.9.2 The two gases of concern for CLD (and HCLD) analysers are CO2 and water vapour. Quench responses of these gases are proportional to their concentrations, and therefore require test techniques to determine the quench at the highest expected concentrations experienced during testing.
CO2 quench check
1.9.2.1 A CO2 span gas having a concentration of 80 to 100% of full scale of the maximum operating range shall be passed through the NDIR analyser and the CO2 value recorded as A. It shall then be diluted approximately 50% with NO span gas and passed through the NDIR and (H)CLD with the CO2 and NO values recorded as B and C, respectively. The CO2 shall be shut off and only the NO span gas be passed through the (H)CLD and the NO value recorded as D.
The quench shall be calculated as follows:
% CO2quench=[1-((C×A)(D×A)-(D×B))]×100
and must not be greater than 3% of full scale,
where:
A:
undiluted CO2 concentration measured with NDIR%
B:
diluted CO2 concentration measured with NDIR%
C:
diluted NO concentration measured with CLD ppm
D:
undiluted NO concentration measured with CLD ppm
Water quench check
1.9.2.2 This check applies to wet gas concentration measurements only. Calculation of water quench must consider dilution of the NO span gas with water vapour and scaling of water vapour concentration of the mixture to that expected during testing. A NO span gas having a concentration of 80 to 100% of full scale to the normal operating range shall be passed through the (H)CLD and the NO value recorded as D. The NO gas shall be bubbled through water at room temperature and passed through the (H)CLD and the NO value recorded as C. The analyser’s absolute operating pressure and the water temperature shall be determined and recorded as E and F, respectively. The mixture’s saturation vapour pressure that corresponds to the bubbler water temperature (F) shall be determined and recorded as G. The water vapour concentration (in %) of the mixture shall be calculated as follows:
H=100× (GE)
and recorded as H. The expected diluted NO span gas (in water vapour) concentration shall be calculated as follows:
De=D× (1-H100)
and recorded as De. For diesel exhaust, the maximum exhaust water vapour concentration (in %) expected during testing shall be estimated, under the assumption of a fuel atom H/C ratio of 1.8 to 1, from the undiluted CO2 span gas concentration (A, as measured in section 1.9.2.1) as follows:
Hm=0.9× A
and recorded as Hm.
The water quench shall be calculated as follows:
% H2OQuench=100×(De-CDe)×(HmH)
and must not be greater than 3% of full scale
De:
expected diluted NO concentration (ppm)
C:
diluted NO concentration (ppm)
Hm:
maximum water vapour concentration (%)
H:
actual water vapour concentration (%)
Note: It is important that the NO span gas contains minimal NO2 concentration for this check, since absorption of NO2 in water has not been accounted for in the quench calculations.
Calibration intervals
1.10 The analysers shall be calibrated according to section 1.5 at least every three months or whenever a system repair or change is made that could influence calibration.
2 CALIBRATION OF THE PARTICULATE MEASURING SYSTEM
Introduction
2.1 Each component shall be calibrated as often as necessary to fulfil the accuracy requirements of this standard. The calibration method to be used is described in this section for the components indicated in Annex III, Apendix 1, section 1.5 and Annex V.
Flow measurement
2.2 The calibration of gas flow-meters or flow measurement instrumentation shall be traceable to national and/or international standards.
The maximum error of the measured value shall be within ± 2% of reading.
If the gas flow is determined by differential measurement, the maximum error of the difference shall be such that the accuracy of GEDFis within ± 4% (see also Annex V, section 1.2.1.1. EGA). It can be calculated by taking the root-mean-square of the errors of each instrument.
Checking the dilution ratio
2.3 When using particulate sampling systems without EGA (Annex V, section 1.2.1.1.), the dilution ratio shall be checked for each new engine installation with the engine running and the use of either the CO2 or NOX concentration measurements in the raw and dilute exhaust.
The measured dilution ratio shall be within ± 10% of the calculated dilution ratio from CO2 or NOX concentration measurement.
Checking the partial flow conditions
2.4 The range of the exhaust gas velocity and the pressure oscillations shall be checked and adjusted according to the requirements of Annex V, section 1.2.1.1, EP, if applicable.
Calibration intervals
2.5 The flow measurement instrumentation shall be calibrated at least every three months, or whenever a system change is made that could influence calibration.

---