# The Weighing Equipment (Automatic Catchweighing Instruments) Regulations 2003, Paragraph


A zero-tracking device shall operate only when:
• the indication is at zero, or at a negative net value equivalent to gross zero, and
• the stability criteria are fulfilled, and
• the corrections are not more than 0.5 e/second.

---