# The Weighing Equipment (Automatic Gravimetric Filling Instruments) Regulations 2000, Schedule 2: (Requirements relating to use)

Extent: E+W+S

Instruments shall comply with the appropriate metrological and technical requirements at temperatures from −10°C to +40°C. However, for special applications the limits of the temperature range may differ from those given above but such a range shall not be less than 30°C and shall be specified in the descriptive markings as set out in Schedule 4.
Where a filling instrument is marked with a temperature range, it shall not be used for trade in temperatures outside that range.

---