# Weighing Equipment (Filling and Discontinuous Totalising Automatic Weighing Machines) Regulations (Northern Ireland) 1986, Paragraph 8


(1) The inspector shall use the subdivided scale provided in accordance with Regulation 24(b) to obtain an accuracy determination to one-tenth of a scale interval at not less than six positions of the weighing range, including zero, minimum load and maximum capacity.
(2) The inspector shall then prepare a calibration chart or graph from the results obtained.

---