Regulation 7 Reference meters
A reference meter shall have been tested at a single flowrate within 6 months before use, and the mean of five consecutive measurements at the same flowrate shall not differ by more than 0.05 per cent. of the quantity delivered on each test from the mean quantity delivered at the same flow rate with liquid of the same viscosity when the meter was last tested in accordance with Regulation 6 above.