Division 3.1.1.5 Calibration curves

United Kingdom

Original text. Later amendments are not shown.

When calibration curves are used for quantification:

  • at least five levels (including zero) should be used in the construction of the curve,

  • the working range of the curve should be described,

  • the mathematical formula of the curve and the goodness-of-fit of the data to the curve should be described,

  • acceptability ranges for the parameters of the curve should be described.

When serial calibration based on a standard solution is necessary, acceptable ranges shall be indicated for the parameters of the calibration curve, which may vary from series to series.