Division 2.3.4 F1F2F3
United Kingdom
In carrying out the complement fixation test, the following controls must be made each time:
- (a)
control of the anti-complementary effect of the serum;
- (b)
control of the antigen;
- (c)
control of sensitised red blood cells;
- (d)
control of the complement;
- (e)
control using a positive serum of sensitivity at the start of the reaction;
- (f)
control of the specificity of the reaction using a negative serum.