Division 7 GAS CHROMATOGRAPHY

United Kingdom

Original text. Later amendments are not shown.

The column must yield a resolution ‘R’ equal to, or better than, 1,5, where:

let: R = 2 d' r 2 - r 1 W 1 + W 2

r1 and r2
  • retention times (in minutes),

  • W1 and W2
  • peak widths at half height (in millimetres),

  • d'
  • the chart speed (in millimetres per minute).

  • The following gas chromatographic conditions have been found suitable:

    Column
  • stainless steel.

  • Length
  • 1,7 m.

  • Diameter
  • 3 mm.

  • Support:

    chromosorb
  • WAW

  • sieve analysis
  • 80 to 100 mesh.

  • Stationary phase
  • 10 % OV 17.

  • Temperatures:

    column
  • 280 oC,

  • injector
  • 280 oC,

  • detector
  • 280 oC.

  • Carrier gas: oxygen-free nitrogen.

    Pressure
  • 2,3 bar.

  • Flow
  • 30 ml/min.