The manufacturer shall specify the rated operating conditions for the instrument, in particular:

  1. 1..

    F1The flowrate range of the water.

    The values for the flowrate range shall fulfil the following conditions:

    • Q 3 /Q 1 ≥ 40

    • Q 2 /Q 1 = 1,6

    • Q 4 /Q 3 = 1,25

  2. 2..

    The temperature range of the water.

    The values for the temperature range shall fulfil the following conditions:

    • 0,1 °C to at least 30 °C, or

    • 30 °C to at least 90 °C.

    The meter may be designed to operate over both ranges.

  3. 3..

    The relative pressure range of the water, the range being 0,3 bar to at least 10 bar at Q3.

  4. 4..

    For the power supply: the nominal value of the AC voltage supply and/or the limits of DC supply.