For minimum measured quantities greater than or equal to 2 litres the following conditions apply:

  • Condition 1

    Emin shall fulfil the condition: Emin ≥ 2 R, where R is the smallest scale interval of the indication device.

  • Condition 2

    Emin is given by the formula: E min = 2MMQ × A ∕ 100 , where:

    • MMQ is the minimum measured quantity,

    • A is the numerical value specified in line A of Table 2.