Division (3) Annex V to Directive 2000/54/EC is replaced by the following:
United Kingdom
‘ANNEX V
INDICATIONS CONCERNING CONTAINMENT MEASURES AND CONTAINMENT LEVELS
(Articles 15(3) and 16(1)(a) and (b))
Preliminary note
The measures contained in this Annex shall be applied according to the nature of the activities, the assessment of risk to workers, and the nature of the biological agent concerned.
In the table, “Recommended” means that the measures should in principle be applied, unless the results of the assessment referred to in Article 3(2) indicate otherwise.
HEPA: High efficiency particulate air
Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.
A. Containment measures B. Containment levels 2 3 4 Workplace 1. The workplace is to be separated from any other activities in the same building No Recommended Yes 2. The workplace is to be sealable to permit fumigation No Recommended Yes Facilities 3. Infected material including any animal is to be handled in a safety cabinet or isolation or other suitable containment Where appropriate Yes, where infection is by airborne route Yes Equipment 4. Input air and extract air to the workplace are to be filtered using (HEPAf1) or likewise No Yes, on extract air Yes, on input and extract air 5. The workplace is to be maintained at an air pressure negative to atmosphere No Recommended Yes 6. Surfaces impervious to water and easy to clean Yes, for bench and floor Yes, for bench, floor and other surfaces determined by risk assessment Yes, for bench, walls, floor and ceiling 7. Surfaces resistant to acids, alkalis, solvents, disinfectants Recommended Yes Yes System of work 8. Access is to be restricted to nominated workers only Recommended Yes Yes, via airlockf2 9. Efficient vector control, for example rodents and insects Recommended Yes Yes 10. Specified disinfection procedures Yes Yes Yes 11. Safe storage of a biological agent Yes Yes Yes, secure storage 12. Personnel should shower before leaving the contained area No Recommended Recommended Waste 13. Validated inactivation process for the safe disposal of animal carcases Recommended Yes, on or off site Yes, on site Other measures 14. A laboratory is to contain its own equipment No Recommended Yes 15. An observation window, or, alternative, is to be present, so that occupants can be seen Recommended Recommended Yes ’
Annotations
Footnotes
- f1
-
HEPA: High efficiency particulate air
- f2
-
Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.