Division (4) Annex VI to Directive 2000/54/EC is replaced by the following:

United Kingdom

Original text. Later amendments are not shown.

‘ANNEX VI

CONTAINMENT FOR INDUSTRIAL PROCESSES

(Article 4(1) and Article 16(2)(a))

Preliminary note

In the table, “Recommended” means that the measures should in principle be applied, unless the results of the assessment referred to in Article 3(2) indicate otherwise.

Group 1 biological agents

For work with group 1 biological agents including live attenuated vaccines, the principles of good occupational safety and hygiene should be observed.

Groups 2, 3 and 4 biological agents

It may be appropriate to select and combine containment requirements from different categories below on the basis of a risk assessment related to any particular process or part of a process.

HEPA: High efficiency particulate air

Closed system: A system that physically separates the process from the environment (e.g. incubator vats, tanks, etc.).

Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.

A. Containment measuresB. Containment levels
234
General
1. Viable organisms should be handled in a system which physically separates the process from the environmentYesYesYes
2. Exhaust gases from the closed system should be treated so as to:Minimise releasePrevent releasePrevent release
3. Sample collection, addition of materials to a closed system and transfer of viable organisms to another closed system, should be performed so as to:Minimise releasePrevent releasePrevent release
4. Bulk culture fluids should not be removed from the closed system unless the viable organisms have been:Inactivated by validated chemical or physical meansInactivated by validated chemical or physical meansInactivated by validated chemical or physical means
5. Seals should be designed so as to:Minimise releasePrevent releasePrevent release
6. The controlled area should be designed to contain spillage of the entire contents of the closed systemNoRecommendedYes
7. The controlled area should be sealable to permit fumigationNoRecommendedYes
Facilities
8. Decontamination and washing facilities should be provided for personnelYesYesYes
Equipment
9. Input air and extract air to the controlled area should be HEPAf1 filteredNoRecommendedYes
10. The controlled area should be maintained at an air pressure negative to atmosphereNoRecommendedYes
11. The controlled area should be adequately ventilated to minimise air contaminationRecommendedRecommendedYes
System of work
12. Closed systemsf2 should be located within a controlled areaRecommendedRecommendedYes, and purpose-built
13. Biohazard signs should be postedRecommendedYesYes
14. Access should be restricted to nominated personnel onlyRecommendedYesYes, via an airlockf3
15. Personnel should shower before leaving the controlled areaNoRecommendedYes
16. Personnel should wear protective clothingYes, work clothingYesYes, complete change
Waste
17. Effluent from sinks and showers should be collected and inactivated before releaseNoRecommendedYes
18. Effluent treatment before final dischargeInactivated by validated chemical or physical meansInactivated by validated chemical or physical meansInactivated by validated chemical or physical means

’

Annotations

Footnotes

f1

HEPA: High efficiency particulate air

f2

Closed system: A system that physically separates the process from the environment (e.g. incubator vats, tanks, etc.).

f3

Airlock: Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.