Division B.3 Apparatus and equipment
United Kingdom
Usual laboratory apparatus and in particular the following:
- B.3.1.
Digital display densimeter
Electronic densimeter for performing such measurements must be capable of expressing density in g/ml to 5 decimal places.
Note 1:
The densimeter should be placed on a perfectly stable stand that is insulated from all vibrations.
- B.3.2.
Temperature regulation
The densimeter's performance is valid only if the measuring cell is connected to a built-in temperature regulator that can achieve the same temperature stability of ± 0,02 °C or better.
Note 2:
The precise setting and monitoring of the temperature in the measuring cell are very important, for an error of 0,1 °C can lead to a variation in density of the order of 0,1 kg/m3.
- B.3.3.
Sample injection syringes or auto sampler.