OpenLegislation

Division B.3 Apparatus and equipment

United Kingdom

Original text. Later amendments are not shown.

Usual laboratory apparatus and in particular the following:

  1. B.3.1.

    Digital display densimeter

    Electronic densimeter for performing such measurements must be capable of expressing density in g/ml to 5 decimal places.

    Note 1:

    The densimeter should be placed on a perfectly stable stand that is insulated from all vibrations.

  2. B.3.2.

    Temperature regulation

    The densimeter's performance is valid only if the measuring cell is connected to a built-in temperature regulator that can achieve the same temperature stability of ± 0,02 °C or better.

    Note 2:

    The precise setting and monitoring of the temperature in the measuring cell are very important, for an error of 0,1 °C can lead to a variation in density of the order of 0,1 kg/m3.

  3. B.3.3.

    Sample injection syringes or auto sampler.