Division B.4 Procedure

United Kingdom

Original text. Later amendments are not shown.

  1. B.4.1.

    Calibration of the densimeter

    The apparatus must be calibrated according to the instrument manufacturer's instructions when it is first put into service. It must be recalibrated regularly and checked against a certified reference standard or an internal laboratory reference solution based on a certified reference standard.

  2. B.4.2.

    Determination of sample density

    1. B.4.2.1.

      If required prior to measurement clean and dry the cell with acetone or absolute alcohol and dry air. Rinse the cell with the sample.

    2. B.4.2.2.

      Inject the sample into the cell (using a syringe or autosampler) so that the cell is completely filled. During the filling operation make sure that all air bubbles are completely eliminated. The sample must be homogeneous and must not contain any solid particles. Any suspended matter should be removed by filtration prior to analysis.

    3. B.4.2.3.

      Once the reading has stabilised, record the density ρ20 or the alcoholic strength displayed by the densimeter.

  3. B.4.3.

    Result

    When the density ρ20 is used, calculate the real alcoholic strength using the alcoholic strength tables identified below:

    The table giving the value of the alcoholic strength by volume (% vol) at 20 °C as a function of the density at 20 °C of water-alcohol mixtures is the international table adopted by the International Legal Metrology Organisation in its Recommendation No 22.