Division C.3.3.1 Performance criteriaF1
United Kingdom
Where no specific methods for the determination of contaminants in foodstuffs are prescribed F2..., laboratories may select any validated method of analysis for the respective matrix provided that the selected method meets the specific performance criteria set out in Tables 5, 6 and 7.
It is recommended that fully validated methods (i.e. methods validated by collaborative trial for the respective matrix) are used where appropriate and available. Other suitable validated methods (e.g. in-house validated methods for the respective matrix) may also be used provided that they fulfil the performance criteria set out in Tables 5, 6 and 7.
Where possible, the validation of in-house validated methods shall include a certified reference material.
- (a)
F3Performance criteria for methods of analysis for lead, cadmium, mercury, inorganic tin and inorganic arsenic
Table 5
Parameter Criterion Applicability Foods specified in Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences Repeatability (RSD r ) HORRAT r less than 2 Reproducibility (RSD R ) HORRAT R less than 2 Recovery The provisions of point D.1.2 apply LOD = three tenths of LOQ LOQ Inorganic tin ≤ 10 mg/kg Lead ML ≤ 0,01 mg/kg 0,01 < ML ≤ 0,02 mg/kg 0,02 < ML < 0,1 mg/kg ML ≥ 0,1 mg/kg ≤ ML ≤ two thirds of the ML ≤ two fifths of the ML ≤ one fifth of the ML Cadmium, mercury, inorganic arsenic ML is < 0,100 mg/kg ML is ≥ 0,100 mg/kg ≤ two fifths of the ML ≤ one fifth of the ML - (b)
F4Performance criteria for methods of analysis for 3-monochloropropane-1,2-diol (3-MCPD), 3-MCPD fatty acid esters and glycidyl fatty acid esters:
Performance criteria for methods of analysis for 3-MCPD in foods specified in point 4.1 of the Annex to Regulation (EC) No 1881/2006
Table 6a
Parameter Criterion Applicability Foods specified in point 4.1 of the Annex to Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences Field blanks Less than LOD Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 75-110 % Limit of Detection (LOD) ≤ 5 μg/kg (on dry matter basis) Limit of Quantification (LOQ) ≤ 10 μg/kg (on dry matter basis) Performance criteria for methods of analysis for 3-MCPD in foods specified in point 4.3 of the Annex to Regulation (EC) No 1881/2006
Table 6b
Parameter Criterion Applicability Foods specified in point 4.3 of the Annex to Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences Field blanks Less than LOD Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 75-110 % Limit of Detection (LOD) ≤ 7 μg/kg Limit of Quantification (LOQ) ≤ 14 μg/kg Performance criteria for methods of analysis for 3-MCPD fatty acid esters, expressed as 3-MCPD , in foods specified in point 4.3 of the Annex to Regulation (EC) No 1881/2006
Table 6c
Parameter Criterion Applicability Foods specified in point 4.3 of the Annex to Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 70-125 % Limit of Detection (LOD) Three tenths of LOQ Limit of Quantification (LOQ)
for foods specified in 4.3.1 and 4.3.2
≤ 100 μg/kg in oils and fats Limit of Quantification (LOQ)
for foods specified in 4.3.3 and in 4.3.4 with a fat content < 40 %
≤ two fifths of the ML Limit of Quantification (LOQ)
for foods specified in 4.3.4 with a fat content ≥ 40 %
≤ 15 μg/kg fat Performance criteria for methods of analysis for glycidyl fatty acid esters, expressed as glycidol, in foods specified in point 4.2 of the Annex to Regulation (EC) No 1881/2006
Table 6d
Parameter Criterion Applicability Foods specified in point 4.2 of the Annex to Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 70-125 % Limit of Detection (LOD) Three tenths of LOQ Limit of Quantification (LOQ)
for foods specified in 4.2.1 and 4.2.2
≤ 100 μg/kg in oils and fats Limit of Quantification (LOQ)
for foods specified in 4.2.3 with a fat content < 65 % and in 4.2.4 with a fat content < 8 %
≤ two fifths of the ML Limit of Quantification (LOQ)
for foods specified in 4.2.3 with a fat content ≥ 65 % and in 4.2.4 with a fat content ≥ 8 %
≤ 31 μg/kg fat
- (c)
Performance criteria for methods of analysis for polycyclic aromatic hydrocarbons:
The four polycyclic aromatic hydrocarbons to which these criteria apply are benzo(a)pyrene, benz(a)anthracene, benzo(b)fluoranthene and chrysene.
Table 7
Parameter Criterion Applicability Foods specified in Regulation (EC) No 1881/2006 Specificity Free from matrix or spectral interferences, verification of positive detection Repeatability (RSD r ) HORRAT r less than 2 Reproducibility (RSD R ) HORRAT R less than 2 Recovery 50-120 % LOD ≤ 0,30 μg/kg for each of the four substances LOQ ≤ 0,90 μg/kg for each of the four substances - (d)
Performance criteria for methods of analysis for acrylamide:
Table 8
Parameter Criterion Applicability All foods Specificity Free from matrix or spectral interferences Field blanks Less than Limit of Detection (LOD) Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 75-110 % Limit of Detection (LOD) Three tenths of LOQ Limit of Quantification (LOQ) For foods with benchmark levels < 125 μg/kg: ≤ two fifths of the benchmark level, however not required to be lower than 20 μg/kg
For foods with benchmark level ≥ 125 μg/kg: ≤ 50 μg/kg
- (e)
F5Performance criteria for methods of analysis for perchlorate:
Table 9
Parameter Criterion Applicability All foods Specificity Free from matrix or spectral interferences Repeatability (RSD r ) 0,66 times RSD R as derived from (modified) Horwitz equation Reproducibility (RSD R ) as derived from (modified) Horwitz equation Recovery 70-110 % Limit of Detection (LOD) Three tenths of LOQ Limit of Quantification (LOQ) ≤ two fifths of the ML - (f)
Notes to the performance criteria:
The Horwitz equationf1 (for concentrations 1,2 x 10 -7 ≤ C ≤ 0,138) and the modified Horwitz equationf2 (for concentrations C < 1,2 × 10 -7 ) are generalised precision equations which are independent of analyte and matrix but solely dependent on concentration for most routine methods of analysis.
Modified Horwitz equation for concentrations C < 1,2 × 10 -7 :
RSD R = 22 %
where:
RSD R is the relative standard deviation calculated from results generated under reproducibility conditions
C is the concentration ratio (i.e. 1 = 100g/100g, 0,001 = 1 000 mg/kg). The modified Horwitz equation applies to concentrations C < 1,2 × 10 -7 .
Horwitz equation for concentrations 1,2 x 10 -7 ≤ C ≤ 0,138:
RSD R = 2C (- 0,15 )
where:
RSD R is the relative standard deviation calculated from results generated under reproducibility conditions
C is the concentration ratio (i.e. 1 = 100g/100g, 0,001 = 1 000 mg/kg). The Horwitz equation applies to concentrations 1,2 x 10 -7 ≤ C ≤ 0,138.
Annotations
Textual amendments
- F1
- F2
Words in Annex para. C.3.3.1 omitted (31.12.2020) by virtue of S.I. 2019/639, reg. 30(e) (as substituted by The Food and Feed Hygiene and Safety (Miscellaneous Amendments etc.) (EU Exit) Regulations 2020 (S.I. 2020/1504), regs. 1(2), 8(10))
- F3
- F4
- F5
Footnotes
- f1
-
W. Horwitz, L.R. Kamps, K.W. Boyer, J.Assoc.Off.Analy.Chem.,63, 1980, 1344-1354.
- f2
-
M. Thompson, Analyst, 125, 2000, 385-386.