OpenLegislation

Division HPLC conditions

United Kingdom

Original text. Later amendments are not shown.

The following conditions are offered for guidance; other conditions may be used provided that they give equivalent results.

Liquid chromatographic column (4.5.1):250 mm × 4 mm, C18, 5 or 10 μm packing, or equivalent
Mobile phase (3.9):Mixture of methanol (3.3) and water e.g. 980 + 20 (v + v).
Flow rate:1-2 ml/min.
Detector (4.5.2):

UV detector (325 nm) or fluorescence detector

(excitation: 325 nm/emission: 475 nm)