Division 5.3.1 Parameters
United Kingdom
The following conditions are offered for guidance, other conditions may be used provided that they give equivalent results.
| Liquid chromatographic column (4.2.1) | 100 mm × 4,6 mm, Hypersil ODS, 3 μm packing, or equivalent | |
| Mobile phase: | Eluent A (3.13.1): | Aqueous solution of ammonium acetate and tetrabutyl-ammonium hydrogen sulphate |
| Eluent B (3.13.2): | acetonitrile | |
| Eluent C (3.13.3): | methanol | |
| Elution mode: |
Flush with B during 10 min. | |
| Flow rate: | 1,5-2 ml/min. | |
| Injection volume: | 20 μl | |
| Detector wavelength: | 280 nm. | |
Check the stability of the chromatographic system, injecting several times the calibration solution (3.10), containing 2 μg/ml, until constant peak heights and retention times are achieved.