Division 5.3.1 Parameters:

United Kingdom

Original text. Later amendments are not shown.

The following conditions are offered for guidance, other conditions may be used provided that they give equivalent results.

Analytical column (4.3.1)
Mobile Phase (3.4):water (3.3)-methanol (3.2) mixture, 900 + 100 (V + V)
Flow rate:1,5-2 ml/min.
Detection wavelength:380 nm
Injection volume:20 μl –100 μl

Check the stability of the chromatographic system, injecting several times the calibration solution (3.5.3) containing 2,5 μg/ml, until constant peak heights and retention times are achieved.