Division 1.5.7.3 Procedure
United Kingdom
The flow rate of the effluent carrier gas is measured at room temperature. The flow rate is checked frequently during the experiment to assure that there is an accurate value for the total volume of carrier gas. Continuous monitoring with a mass flow-meter is preferred. Saturation of the gas phase may require considerable contact time and hence quite low gas flow rates (25).
At the end of the experiment, both the front and backup sorbent sections are analysed separately. The compound on each section is desorbed by adding a solvent. The resulting solutions are analysed quantitatively to determine the weight desorbed from each section. The choice of the analytical method (also the choice of sorbent and desorbing solvent) is dictated by the nature of the test material. The desorption efficiency is determined by injecting a known amount of sample onto the sorbent, desorbing it and analysing the amount recovered. It is important to check the desorption efficiency at or near the concentration of the sample under the test conditions.
To assure that the carrier gas is saturated with the test substance, three different gas flow rates are used. If the calculated vapour pressure shows no dependence on flow rate, the gas is assumed to be saturated.
The vapour pressure is calculated through the equation:
where:
vapour pressure (Pa)
mass of evaporated test substance (g)
volume of saturated gas (m3)
universal gas constant 8,314 (J mol–1 K–1)
temperature (K)
molar mass of test substance (g mol–1)
Measured volumes must be corrected for pressure and temperature differences between the flow meter and the saturator.