Division 4.2.2.2 Linear response of an analytical balanceF1
The analytical balance used to determine the filter weight shall meet the linearity verification criteria of Table A5/1 applying a linear regression. This implies a precision of at least ± 2 μg and a resolution of at least 1 μg (1 digit = 1 μg). At least 4 equally-spaced reference weights shall be tested. The zero value shall be within ± 1 μg.
Table A5/1
Analytical balance verification criteria
| Measurement system | Intercept a0 | Slope a1 | Standard error of estimate (SEE) | Coefficient of determination r2 |
|---|---|---|---|---|
| Particulate balance | ≤ 1 μg | 0,99 – 1,01 | ≤ 1 per cent max | ≥ 0,998 |