Division 2.10.1.2.3 F1
It is recommended that the microbalance be checked at the start of each weighing session, within 24 hours of the sample weighing, by weighing one reference item of approximately 100 mg. This item shall be weighed three times and the arithmetic average result included in all relevant test sheets. If the arithmetic average result of the weighings is ± 5 μg of the result from the previous weighing session, the weighing session and balance are considered valid.