Division 2.11.1.2.2 F1
Each day, a zero check on the PNC, using a filter of appropriate performance at the PNC inlet, shall report a concentration of ≤ 0,2 particles per cm3. Upon removal of the filter, the PNC shall show an increase in measured concentration to at least 100 particles per cm3 when sampling ambient air and a return to ≤ 0,2 particles per cm3 on replacement of the filter.