Division 3 The method of analysis used for the analysis of acrylamide must comply with the following performance criteria

United Kingdom

Original text. Later amendments are not shown.

ParameterCriterion
ApplicabilityFoods specified in this Regulation
SpecificityFree from matrix or spectral interferences
Field blanksLess than Limit of Detection (LOD)
Repeatability (RSDr)0,66 times RSDR as derived from (modified) Horwitz equation
Reproducibility (RSDR)as derived from (modified) Horwitz equation
Recovery75-110 %
Limit of Detection (LOD)Three tenths of LOQ
Limit of Quantification (LOQ)

For benchmark level < 125 μg/kg: ≤ two fifths of the benchmark level (however not required to be lower than 20 μg/kg)

For benchmark level ≥ 125 μg/kg: ≤ 50 μg/kg