Division 8.1.11.1.2 Measurement principles
United Kingdom
H2O and CO2 can negatively interfere with a CLD's NOx response by collisional quenching, which inhibits the chemiluminescent reaction that a CLD utilizes to detect NOx. This procedure and the calculations in point 8.1.11.2.3 determine quench and scale the quench results to the maximum mole fraction of H2O and the maximum CO2 concentration expected during emission testing. If the CLD analyzer uses quench compensation algorithms that utilize H2O and/or CO2 measurement instruments, quench shall be evaluated with these instruments active and with the compensation algorithms applied.