Division 8.1.11.1.3 System requirements

United Kingdom

Original text. Later amendments are not shown.

For dilute measurement a CLD analyzer shall not exceed a combined H2O and CO2 quench of ± 2 %. For raw measurement a CLD analyzer shall not exceed a combined H2O and CO2 quench of ± 2,5 %. Combined quench is the sum of the CO2 quench determined as described in point 8.1.11.1.4 and the H2O quench as determined in point 8.1.11.1.5. If these requirements are not met, corrective action shall be taken by repairing or replacing the analyzer. Before running emission tests, it shall be verified that the corrective action have successfully restored the analyzer to proper functioning.