OpenLegislation

Division 8.1.8.4.1 Overview

United Kingdom

Original text. Later amendments are not shown.

  1. (a)

    This section describes how to calibrate flow meters for diluted exhaust gas constant-volume sampling (CVS) systems;

  2. (b)

    This calibration shall be performed while the flow meter is installed in its permanent position. This calibration shall be performed after any part of the flow configuration upstream or downstream of the flow meter has been changed that may affect the flow-meter calibration. This calibration shall be performed upon initial CVS installation and whenever corrective action does not resolve a failure to meet the diluted exhaust gas flow verification (i.e., propane check) in point 8.1.8.5;

  3. (c)

    A CVS flow meter shall be calibrated using a reference flow meter such as a subsonic venturi flow meter, a long-radius flow nozzle, a smooth approach orifice, a laminar flow element, a set of critical flow venturis, or an ultrasonic flow meter. A reference flow meter shall be used that reports quantities that are internationally-traceable within ± 1 % uncertainty. This reference flow meter's response to flow shall be used as the reference value for CVS flow-meter calibration;

  4. (d)

    An upstream screen or other pressure restriction that could affect the flow ahead of the reference flow meter may not be used, unless the flow meter has been calibrated with such a pressure restriction;

  5. (e)

    The calibration sequence described under this point 8.1.8.4 refers to the molar based approach. For the corresponding sequence used in the mass based approach, see point 2.5 of Annex VII.

  6. (f)

    F1CFV or SSV may alternatively be removed from its permanent position for calibration as long as the following requirements are met when installed in the CVS:

    1. (1)

      Upon installation of the CFV or SSV into the CVS, good engineering judgment shall be applied to verify that any leaks have not been introduced between the CVS inlet and the venturi.

    2. (2)

      After ex-situ venturi calibration, all venturi flow combinations must be verified for CFVs or at minimum of 10 flow points for an SSV using the propane check as described in point 8.1.8.5. The result of the propane check for each venturi flow point may not exceed the tolerance in point 8.1.8.5.6.

    3. (3)

      In order to verify the ex-situ calibration for a CVS with more than a single CFV, the following verification shall be conducted:

      1. (i)

        A constant flow device shall be used to deliver a constant flow of propane to the dilution tunnel.

      2. (ii)

        The hydrocarbon concentrations shall be measured at a minimum of 10 separate flow rates for an SSV flow meter, or at all possible flow combinations for a CFV flow meter, while keeping the flow of propane constant.

      3. (iii)

        The concentration of hydrocarbon background in the dilution air shall be measured at the beginning and end of this test. The average background concentration from each measurement at each flow point must be subtracted before performing the regression analysis in paragraph (iv).

      4. (iv)

        A power regression has to be performed using all the paired values of flow rate and corrected concentration to obtain a relationship in the form of y = a × xb, using the concentration as the independent variable and the flow rate as the dependent variable. For each data point, the calculation of the difference between the measured flow rate and the value represented by the curve fit is required. The difference at each point must be less than ± 1 % of the appropriate regression value. The value of b must be between – 1,005 and – 0,995. If the results do not meet these limits, corrective actions consistent with point 8.1.8.5.1(a) must be taken.

Annotations

Textual amendments

F1

Substituted by Commission Delegated Regulation (EU) 2018/989 of 18 May 2018 amending and correcting Delegated Regulation (EU) 2017/654 supplementing Regulation (EU) 2016/1628 of the European Parliament and of the Council with regard to technical and general requirements relating to emission limits and type-approval for internal combustion engines for non-road mobile machinery (Text with EEA relevance).