Division 8.1.8.7.4 Dilution-of-span-gas leak test
United Kingdom
Any gas analyzer may be used for this test. If a FID is used for this test, any HC contamination in the sampling system shall be corrected in accordance with sections 2 or 3 of Annex VII on HC determination. Misleading results shall be avoided by using only analyzers that have a repeatability of 0,5 % or better at the span gas concentration used for this test. The vacuum side leak check shall be performed as follows:
- (a)
A gas analyzer shall be prepared as it would be for emission testing;
- (b)
Span gas shall be supplied to the analyzer port and it shall be verified that the span gas concentration is measured within its expected measurement accuracy and repeatability;
- (c)
Overflow span gas shall be routed to one of the following locations in the sampling system:
- (i)
The end of the sample probe;
- (ii)
The transfer line shall be disconnected at the probe connection, and the span gas overflown at the open end of the transfer line;
- (iii)
A three-way valve installed in-line between a probe and its transfer line;
- (i)
- (d)
It shall be verified that the measured overflow span gas concentration is within ± 0,5 % of the span gas concentration. A measured value lower than expected indicates a leak, but a value higher than expected may indicate a problem with the span gas or the analyzer itself. A measured value higher than expected does not indicate a leak.