Division 8.1.8.7.3 Low-flow leak test

United Kingdom

Original text. Later amendments are not shown.

A sampling system shall be tested for low-flow leaks as follows:

  1. (a)

    The probe end of the system shall be sealed by taking one of the following steps:

    1. (i)

      The end of the sample probe shall be capped or plugged;

    2. (ii)

      The transfer line shall be disconnected at the probe and the transfer line capped or plugged;

    3. (iii)

      A leak-tight valve in-line between a probe and transfer line shall be closed;

  2. (b)

    All vacuum pumps shall be operated. After stabilizing, it shall be verified that the flow through the vacuum-side of the sampling system is less than 0,5 % of the system's normal in-use flow rate. Typical analyzer and bypass flows may be estimated as an approximation of the system's normal in-use flow rate.