Regulation 13 Procedures for affixing a UK marking to general medical devicesF8E1
England, Wales and Scotland
A relevant device falling within Class I may bear a F1UK marking only if its manufacturer or F2their UK responsible person—
fulfils the applicable obligations imposed by Annex VII;
declares, in accordance with the declaration of conformity procedure set out in that Annex, that the device meets the provisions of F3this Part which apply to it; and
ensures that the device meets the provisions of this Part which apply to it.
A relevant device falling within Class IIa may bear a UK marking only if its manufacturer or their UK responsible person—
fulfils the applicable obligations imposed by—
Annex II, excluding Section 4 of that Annex, or
Annex VII, together with Annex IV, V or VI;
declares, in accordance with a declaration of conformity procedure set out in those Annexes, that the device meets the provisions of this Part which apply to it; and
ensures that the device meets the provisions of this Part which apply to it.
A relevant device falling within Class IIb may bear a UK marking only if its manufacturer or their UK responsible person—
fulfils the applicable obligations imposed by—
Annex II, excluding Section 4 of that Annex, or
Annex III, together with Annex IV, V or VI;
declares, in accordance with a declaration of conformity procedure set out in those Annexes, that the device meets the provisions of this Part which apply to it; and
ensures that the device meets the provisions of this Part which apply to it.
A relevant device falling within Class III may bear a UK marking only if its manufacturer or their UK responsible person—
fulfils the applicable obligations imposed by—
Annex II, or
Annex III, together with Annex IV or V;
declares, in accordance with a declaration of conformity procedure set out in those Annexes, that the device meets the provisions of this Part which apply to it; F4...
ensures that the device meets the provisions of this Part which apply to it; F5and
fulfils the obligations imposed by Regulation (EU) No 722/2012 (if applicable).
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Annotations
Textual amendments
- F1
Words in reg. 13 substituted (E.W.S.) (31.12.2020) by The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/791), regs. 1(1), 4(7A)(c) (as amended by S.I. 2020/1478, reg. 1(3), Sch. 2 paras. 2, 17); 2020 c. 1, Sch. 5 para. 1(1)
- F2
Words in reg. 13 substituted (E.W.S.) (31.12.2020) by The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/791), regs. 1(1), 4(7A)(d) (as amended by S.I. 2020/1478, reg. 1(3), Sch. 2 paras. 2, 17); 2020 c. 1, Sch. 5 para. 1(1)
- F3
Words in reg. 13 substituted (E.W.S.) (31.12.2020) by The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/791), regs. 1(1), 4(7A)(b) (as amended by S.I. 2020/1478, reg. 1(3), Sch. 2 paras. 2, 17); 2020 c. 1, Sch. 5 para. 1(1)
- F4
Word in reg. 13(4)(b) omitted (21.10.2013) by virtue of The Medical Devices (Amendment) Regulations 2013 (S.I. 2013/2327), regs. 1(2), 4(2)(a)
- F5
Word in reg. 13(4)(c) added (21.10.2013) by The Medical Devices (Amendment) Regulations 2013 (S.I. 2013/2327), regs. 1(2), 4(2)(b)
- F6
Reg. 13(4)(d) added (21.10.2013) by The Medical Devices (Amendment) Regulations 2013 (S.I. 2013/2327), regs. 1(2), 4(2)(c)
- F7
Reg. 13(5)(6) omitted (21.10.2013) by virtue of The Medical Devices (Amendment) Regulations 2013 (S.I. 2013/2327), regs. 1(2), 4(3)
- F8
Words in reg. 13 heading substituted (E.W.S.) (31.12.2020) by The Medical Devices (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/791), regs. 1(1), 4(7A)(a) (as amended by S.I. 2020/1478, reg. 1(3), Sch. 2 paras. 2, 17); 2020 c. 1, Sch. 5 para. 1(1)
Extent
- E1
This version of this provision extends to England and Wales and Scotland only; a separate version has been created for Northern Ireland only