Paragraph 6 Test loads

Original text. Later amendments are not shown.

(1)

Every meter shall be tested at each of the loads specified in Table 2.

(2)

The ratio accuracy shall be determined for voltage transformers intended for use with meters but are not tested with a meter.

(3)

Current transformers intended for use with meters but not tested with a meter shall be tested from 5% to 100% of rated current.