OpenLegislation

Regulation 10 Passing as fit for use for tradeI1

Northern Ireland

(1)

An inspector shall not pass as fit for use for trade a filling instrument unless –

(a)

it complies with all the appropriate requirements of these Regulations; and

(b)

on testing, it falls within the prescribed limits of error in relation to passing as fit for use for trade.

(2)

An inspector shall not pass as fit for use for trade a filling instrument imported from F1an EEA State unless –

(a)

where the requisite documentation is presented in accordance with regulation 9, the test report recites and the test data confirm to the satisfaction of the inspector that, on testing in accordance with the provisions of clause 5.3 (initial verification) of Part 1 of OIML R 61, that filling instrument fell within limits of error which afford in use an equivalent standard to the prescribed limits of error; and

(b)

it otherwise complies with all the relevant requirements of these Regulations.

(3)

For the purposes of these Regulations and subject to regulation 5(c), the prescribed limits of error relating to a filling instrument shall be determined –

(a)

in accordance with the provisions of Schedule 5; and

(b)

where test fills are required, on the basis of consecutive fills.

Annotations

Textual amendments

F1

Words in reg. 10(2) substituted (31.12.2020) by The Metrology, Health and Safety and Product Safety (Amendment) (Northern Ireland) (EU Exit) Regulations 2019 (S.I. 2019/202), reg. 1(2), Sch. 2 para. 9(3); 2020 c. 1, Sch. 5 para. 1(1)

Commencement

I1

Reg. 10 in operation at 1.3.2005, see reg. 1(1)