Division 5.2.2 Choice of operating conditionsF1
United Kingdom
The operating conditions are generally as follows:
-
column temperature:
20 °C/minute 5 °C/minute 20 °C/minute Initially 80 °C
(1′)
→ 240 °C → 325 °C
(6′)
→ 340 °C
(10′)
-
detector temperature: 350 °C;
-
quantity of substance injected: 1 μl of the n-heptane solution (2-4 ml);
-
carrier gas: helium or hydrogen at the correct linear velocity for the gas selected (see Appendix);
-
instrument sensitivity: suitable for the following conditions:
The conditions may be modified according to the characteristics of the column and the GC apparatus to obtain separation of all the waxes and a satisfactory peak resolution (see figure); the internal standard C 32 retention time must be 18 ± 3 minutes. The most representative wax peak must be at least 60 % of the full scale.
The peak integration parameters must be established so as to obtain a correct evaluation of the areas of the peaks in question.