Division 5.2.1 Preliminary procedureF1
United Kingdom
Fit the column to the gas chromatograph (point 3.3), connecting the inlet port to the on-column system and the outlet port to the detector. Check the gas chromatography apparatus (operation of gas loops, efficiency of detector and recorder system, etc.).
If the column is being used for the first time, it is advisable to condition it. Run a light flow of gas through the column, then switch on the gas chromatography apparatus. Gradually heat until a temperature of 350 °C is reached after approximately 4 h.
Maintain this temperature for at least 2 h, then regulate the apparatus to the operating conditions (regulate gas flow, light flame, connect to electronic recorder (point 3.3.4), regulate oven temperature for column, regulate detector, etc.). Record the signal at a sensitivity at least twice as high as that required for the analysis. The base line should be linear, with no peaks of any kind, and must not have any drift.
Negative straight-line drift indicates that the column connections are not correct while positive drift indicates that the column has not been properly conditioned.
Annotations
Textual amendments
Notes on Annex XX
Footnotes
- f1
-
After elution of the sterol esters, the chromatogram should not show any significant peaks (triacylglycerols).