Division 5.2.2 Choice of operating conditions for waxes and methyl and ethyl esters (Note 6).F1

United Kingdom

Original text. Later amendments are not shown.

The operating conditions are generally as follows:

— Column temperature
  • 20 °C/min 5 °C/min

    80 °C at first (1′) 140 °C 335 °C (20)

  • — Detector temperature
  • 350 °C.

  • — Amount injected
  • 1 μl of n-heptane solution (2-4 ml).

  • — Carrier gas
  • helium or hydrogen at the optimal linear speed for the gas chosen (see Appendix A).

  • — Instrument sensitivity
  • suitable for fulfilling the above conditions.

  • These conditions may be modified to suit the characteristics of the column and the gas chromatograph in order to separate all the waxes and fatty acid methyl and ethyl esters and to obtain satisfactory peak separation (see Figures 2, 3 and 4) and a retention time of 18 ± 3 minutes for the lauryl arachidate internal standard. The most representative peak of the waxes must be over 60 % of the full-scale value while the methyl heptadecanoate internal standard for the methyl and ethyl esters must reach the full-scale value.

    The peak integration parameters should be determined in such a way as to obtain a correct evaluation of the peak areas considered.

    1. Note 6:Division Note 6:

    Annotations

    Textual amendments

    F1

    Inserted by Commission Regulation (EU) No 61/2011 of 24 January 2011 amending Regulation (EEC) No 2568/91 on the characteristics of olive oil and olive-residue oil and on the relevant methods of analysis.

    Notes on Annex XX

    Footnotes

    f1

    After elution of the sterol esters, the chromatogram should not show any significant peaks (triacylglycerols).