Division 5.3.1 Parameters

United Kingdom

Original text. Later amendments are not shown.

The following conditions are offered for guidance; other conditions may be used, provided they yield equivalent results:

Liquid chromatographic column (4.3.1):125 mm × 4 mm, reversed-phase C18, 5 μm packing or equivalent
Mobile phase (3.9):Mixture of phosphate buffer solution (3.7) and methanol (3.5), 5+95 (V+V)
Flow rate:1,2 ml/min.
Detection wavelengths:
Excitation:310 nm
Emission:419 nm
Injection volume:20 μl

Check the stability of the chromatographic system, injecting the calibration solution (3.10.3) containing 4,0 μg/ml several times, until constant peak heights (or areas) and retention times are achieved.