Division 5.3.1 Parameters:

United Kingdom

Original text. Later amendments are not shown.

The following conditions are offered for guidance, other conditions may be used provided that they give equivalent results.

Liquid chromatographic
column (4.1.1):125 mm × 4 mm, cation exchange Nucleosil 10 SA, 5 or 10 μm packing, or equivalent
Mobile phase (3.6):Mixture of acetonitrile (3.2), sodium dihydrogen phosphate solution (3.4) and sodium perchlorate solution (3.5), 450+450+100 (v+v+v).
Flow rate:0,7-1 ml/min
Detection wavelength:264 nm
Injection volume:100 μl

Check the stability of the chromatographic system, injecting several times the calibration solution (3.7.3) containing 1,0 μg/ml, until constant peak heights and retention times are achieved.