HPLC equipment, consisting of:

  1. 6.11.1.

    Pump

  2. 6.11.2.

    Injector, hand or automatic, with a 15 to 30 μl capacity

  3. 6.11.3.

    Two TSK 2 000 -SW columns in series (length 30 cm, internal diameter 0,75 cm) or equivalent columns (e.g. single TSK 2 000 -SWxl, single Agilent Technologies Zorbax GF 250) and a precolumn (3 cm × 0,3 cm) packed with I 125 or material of equivalent effectiveness

  4. 6.11.4.

    Thermostatic column oven, set at 35 ± 1 °C

  5. 6.11.5.

    Variable wavelength UV detector, permitting measurements at 205 nm with a sensitivity of 0,008 Å

  6. 6.11.6.

    Integrator capable of valley-to-valley integration

    Note: Working with columns kept at room temperature is possible, but their power of resolution is slightly lower. In that case, the temperature should vary by less than ± 5 °C in any one range of analyses.