OpenLegislation

Division 6.11 F1

United Kingdom

Original text. Later amendments are not shown.

HPLC equipment, consisting of:

  1. 6.11.1.

    Binary gradient pumping system

  2. 6.11.2.

    Injector, hand or automatic, with a 100 μl capacity

  3. 6.11.3.

    Agilent Technologies Zorbax 300 SB-C3 column (length 25 cm, 0,46 cm internal diameter) or an equivalent wide-pore silica based reversed-phase column

  4. 6.11.4.

    Thermostatic column oven, set at 35 ± 1 °C

  5. 6.11.5.

    Variable wavelength UV detector, permitting measurements at 210 nm (if necessary, a higher wavelength up to 220 nm may be used) with a sensitivity of 0,02 Å

  6. 6.11.6.

    Integrator capable of setting the integration to common baseline or valley-to-valley

    Note : Operation of the column at room temperature is possible, provided that the room temperature does not fluctuate more than 1 °C, otherwise too much variation in the retention time of CMP A takes place.

Annotations

Textual amendments

F1

Inserted by Commission Implementing Regulation (EU) 2018/150 of 30 January 2018 amending Implementing Regulation (EU) 2016/1240 as regards methods for the analysis and quality evaluation of milk and milk products eligible for public intervention and aid for private storage.