Division 4.3.1.3.4 The PNC shall:

Original text. Later amendments are not shown.

  1. (a)

    Operate under full flow operating conditions;

  2. (b)

    Have a counting accuracy of ± 10 per cent across the range 1 per cm3 to the upper threshold of the single particle count mode of the PNC against a suitable traceable standard. At concentrations below 100 per cm3, measurements averaged over extended sampling periods may be required to demonstrate the accuracy of the PNC with a high degree of statistical confidence;

  3. (c)

    Have a resolution of at least 0.1 particles per cm3 at concentrations below 100 per cm3;

  4. (d)

    Have a linear response to particle number concentrations over the full measurement range in single particle count mode;

  5. (e)

    Have a data reporting frequency equal to or greater than a frequency of 0,5 Hz;

  6. (f)

    Have a t90 response time over the measured concentration range of less than 5 seconds;

  7. (g)

    Incorporate a coincidence correction function up to a maximum 10 per cent correction, and may make use of an internal calibration factor as determined in paragraph 5.7.1.3.of this Sub-Annex but shall not make use of any other algorithm to correct for or define the counting efficiency;

  8. (h)

    Have counting efficiencies at the different particle sizes as specified in Table A5/2.

Table A5/2
PNC counting efficiency
Particle size electrical mobility diameter (nm)PNC counting efficiency (per cent)
23 ± 150 ± 12
41 ± 1> 90