Division 3.2.1.1.3.2 F1
For methane measurement using an NMC-FID, the calculation of NMHC depends on the calibration gas/method used for the zero/calibration adjustment.
The FID used for the THC measurement (without NMC) shall be calibrated with propane/air in the normal manner.
For the calibration of the FID in series with an NMC, the following methods are permitted:
- (a)
The calibration gas consisting of propane/air bypasses the NMC;
- (b)
The calibration gas consisting of methane/air passes through the NMC.
It is highly recommended to calibrate the methane FID with methane/air through the NMC.
In case (a), the concentration of CH4 and NMHC shall be calculated using the following equations:
If RfCH4 < 1,05, it may be omitted from the equation above for CCH4.
In case (b), the concentration of CH4 and NMHC shall be calculated using the following equations:
where:
is the HC concentration with sample gas flowing through the NMC, ppm C;
is the HC concentration with sample gas bypassing the NMC, ppm C;
is the methane response factor as determined per paragraph 5.4.3.2. of Sub-Annex 5;
is the methane efficiency as determined per paragraph 3.2.1.1.3.3.1. of this Sub-Annex;
is the ethane efficiency as determined per paragraph 3.2.1.1.3.3.2. of this Sub-Annex.
If RfCH4 < 1,05, it may be omitted in the equations for case (b) above for CCH4 and CNMHC.