Division 3.4.1.2

United Kingdom

Original text. Later amendments are not shown.

The following system requirements shall be met by a single torque meter by calibration:

  1. i..

    Non linearity: < ± 6 Nm

  2. ii..

    Repeatability: < ± 6 Nm

  3. iii..

    Crosstalk: < ± 1 % FSO (only applicable for rim torque meters)

  4. iv..

    Measurement rate: ≥ 20 Hz

where:

  • ‘Non linearity’ means the maximum deviation between ideal and actual output signal characteristics in relation to the measurand in a specific measuring range.

  • ‘Repeatability’ means closeness of the agreement between the results of successive measurements of the same measurand carried out under the same conditions of measurement.

  • ‘Crosstalk’ means signal at the main output of a sensor (My), produced by a measurand (Fz) acting on the sensor, which is different from the measurand assigned to this output. Coordinate system assignment is defined according to ISO 4130.

  • ‘FSO’ means full scale output of calibrated range.

The recorded torque data shall be corrected for the instrument error determined by the supplier.